Quaternary Ammonium Compounds (QAC)

-Direct Binary Complex Method (Method 8337)

Scope:

The reagents buffer the sample to an acidic pH and include a masking agent to remove potential interferences. The indicator reacts with quaternary ammonium compounds in the sample and forms a pale pink to vivid purple color. This test is applicable to monitor quaternary ammonium compounds in swimming pools and cooling towers. The measurement wavelength is 575 nm.

 

Method Data

Method T288-91
28 days
To preserve samples for later analysis, adjust the sample pH to less than 2 with concentrated sulfuric acid (approximately 2 mL per liter). No acid addition is necessary if the sample is tested immediately. Keep the preserved samples at or below 6 °C.
Collect samples in glass bottles that have been rinsed several times with sample. Do not use plastic containers, which can adsorb quaternary ammonium compounds.
Wenner Four Pin
HACH Method 8337

Analyte List

AnalyteFormulaCAS Number
Quaternary Ammonium Compounds

* The typical detection limits and analytes are displayed for this particular method. Please contact Specialty Analytical if other analytes or detection limits are required in addition to the 200.8 Analyte List.

Request A Quote

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we’ll provide you a quote via email. As you are searching or browsing our Analytical Guide, you’ll see the Request-A-Quote icon… just click on it to start the request process.